RIQE: a NIQE-style reference model for Computed Tomography
2610.00384

Authors

Fabio Mattiussi

Abstract

The Natural Image Quality Evaluator (NIQE) scores an image by its statistical distance from a model fitted on pristine images, and its distributed model is fitted on photographs. We release the Radiology Image Quality Evaluator (RIQE), a NIQE-style model fitted on 3,792 full-dose slices from 158 patients of the public LDCT-and-Projection-data collection, with a declared intensity mapping, a manifest of every slice and a script that reproduces the fit.

On 40 held-out patients, RIQE ranks reduced-dose reconstructions, simulated by projection-domain noise insertion, worse than the full-dose reconstruction of the same slice in 240 of 240 chest and 230 of 240 abdominal pairs, and ranks images with 20% more noise worse than their source in 97.5-100% of cases. Its preferences among filtered images, however, do not follow lesion signal.

With a 4 mm, +10 HU lesion inserted in noisy abdominal slices, RIQE prefers bilateral filtering to the unfiltered image in every image up to a 32 HU residual, at which 29% of the lesion's matched-filter signal remains and its detectability index falls from 0.51 to 0.33; it never prefers Gaussian smoothing, which at the same 32 HU residual leaves 70% of the signal and a detectability index of 0.48. Fitted on photographs with the parameters published for NIQE, the same code ranks every simulated reduced-dose abdominal image better than its full-dose counterpart.

RIQE is suited to ranking a degraded image against its source; under the conditions tested it should not be the sole criterion for selecting, comparing or tuning denoisers.

Resources

Ray graphicRay graphicRay graphicRay graphic

Stay in the loop

Every AI paper that matters, free in your inbox daily.

Details

  • takara.ai
  • Custom AI and machine learning from the Frontier Research Team.
  • © 2026 takara.ai Ltd
  • Content is sourced from third-party publications.
Ray graphicRay graphicRay graphic